PERFORMANCE EVALUATION BETWEEN TWO AUTOMATED BIOCHEMICAL ANALYZER SYSTEMS: ROCHE COBAS 8000 AND MINDRAY BS2000M: Evaluated performance of Roche Cobas 8000 and Mindray BS2000M
Scindeks Assistant Scindeks Assistant — A system for serious journals and those aspiring to become one
PDF

Abstract

Summary

Background. The values of biomarkers play a central role in routine clinical decision-making. Whereas, the performances of different automated chemical analyzers remain unclear. To determine the performances of different platforms, we evaluated the capability between Roche Cobas 8000 and Mindray BS2000M. 

Methods. A total of 1869 remaining serum samples were collected. CK, LDH-1, RBP, Cys-c, IgA, IgM and IgG were assessed by using paired-t test, Passing-Bablok regression analysis and Bland Altman analysis according to CLSI EP5-A3.

Results. There were significant in average bias of all items between two machines (P < 0.001). Due to the 95% confidence interval of intercept A included 0, CK, LDH-1, Cys-c and IgG were not show systemic error in Passing-Bablok regression analysis. Except for IgA, the r values and correlation coefficient of all items were higher than 0.91, which showed that the correlation and consistency is good. The Bland-Altman analysis showed that two instruments had more than 95% of the points apart from CK, LDH-1, and IgA.

Conclusions. It can be considered that the two instruments have good correlation and consistency in CK, LDH-1, RBP, Cys-c, IgM and IgG, and the two instruments are interchangeable and can replace each other.

Keywords

Array
Array
Array
Array
Array
DOI: 10.5937/jomb0-34328

References

The published articles will be distributed under the Creative Commons Attribution 4.0 International License (CC BY). It is allowed to copy and redistribute the material in any medium or format, and remix, transform, and build upon it for any purpose, even commercially, as long as appropriate credit is given to the original author(s), a link to the license is provided and it is indicated if changes were made. Users are required to provide full bibliographic description of the original publication (authors, article title, journal title, volume, issue, pages), as well as its DOI code. In electronic publishing, users are also required to link the content with both the original article published in Journal of Medical Biochemistry and the licence used.