Uticaj autoverifikacije na vreme izdavanja rezultata za kliničko-hemijske i imunoesejske testove: Prvo iskustvo sa laboratorijskom autoverifikacijom
Scindeks Asistent Scindeks Asistent — sistem za ozbiljne časopise i one koji to žele da postanu
PDF (engleski)

Sažetak

Background: Auto-verification is increasingly recognized as a key tool for improving quality and efficiency in clinical laboratories. This study aimed to investigate the impact of an auto-verification system implemented for clinical chemistry and immunoassay tests on turnaround time (TAT) in a tertiary-care medical biochemistry laboratory.

Methods: This study was conducted in the Medical Biochemistry Laboratory of XXX Hospital, a tertiary healthcare institution. Clinical chemistry and immunoassay tests were subjected to auto-verification using the navify® Lab Operations middleware (Roche Diagnostics, Germany) integrated with the MIA-MED Laboratory Information System (MIA Technology, Turkey). Algorithms were developed in accordance with CLSI AUTO10-A and AUTO15 guidelines, incorporating rules for quality control, serum indices, analyzer flags, delta checks, critical values, consistency checks, and analytical measurement intervals, as well as recommendations from the national health authorities. Validation of the algorithms was carried out using both simulated and patient data. The proportion of results exceeding predefined TAT targets was compared before and after auto-verification implementation with using the chi-square test. p<0.05 was considered statistically significant.

Results
Overall, 71% of test results and 21% of tube-based results were verified automatically. Median TAT was reduced by 6 minutes for emergency tests and 12 minutes for routine tests. The proportion of results exceeding the TAT threshold decreased significantly from 6.4% before auto-verification to 5.8% after auto-verification implementation (p < 0.001).

Conclusions
Auto-verification, with clearly defined and validated rules, enhances both the reliability and timeliness of laboratory results, thereby supporting quality improvement initiatives in clinical laboratories

Ključne reči

Array
Array
Array
Array
Array
DOI: 10.5937/jomb0-61380

Reference

The published articles will be distributed under the Creative Commons Attribution 4.0 International License (CC BY). It is allowed to copy and redistribute the material in any medium or format, and remix, transform, and build upon it for any purpose, even commercially, as long as appropriate credit is given to the original author(s), a link to the license is provided and it is indicated if changes were made. Users are required to provide full bibliographic description of the original publication (authors, article title, journal title, volume, issue, pages), as well as its DOI code. In electronic publishing, users are also required to link the content with both the original article published in Journal of Medical Biochemistry and the licence used.

Authors are able to enter into separate, additional contractual arrangements for the non-exclusive distribution of the journal's published version of the work (e.g., post it to an institutional repository or publish it in a book), with an acknowledgement of its initial publication in this journal.

Preuzimanja

Podaci o preuzimanju još nisu dostupni.